Product Features | ||
Medical & ITE safety approvals | ||
2MOPP input to output isolation | ||
Leakage current≤100μA | ||
Efficiency level V | ||
Standby power ≤0.3W | ||
3V to 24V outputs, up to 15W | ||
Interchangeable AC plugs | ||
Technical Data | ||
Input Voltage Range:90~264VAC | ||
Input Current: 0.5A at 100-240VAC | ||
Output voltage Tolerance:±5% | ||
Environmental Requirements | ||
Operating Temperature: 0~40℃ | ||
Storage Temperature:-20~60℃ | ||
Operating Humidity:10~90%RH, non-condensing | ||
Storage Humidity: 5~90%RH | ||
Operating Altitude: 2,000 M | ||
Safety Approvals | ||
Safety aAgency/Mark | Medical | I.T.E. |
CB | IEC60601-1 | IEC60950-1 |
UL | UL60601-1 | UL60950-1 |
- | CAN/CSA C22.2 NO.60950-1 | |
TUV-SUD/Mark | EN60601-1 | |
TUV-SUD/GS | - | EN60950-1 |
RCM | - | AS/NZS 60950.1 |
CE | EN60950 | |
CCC | GB4943.1 | |
PSE | J60950 | |
KC | K60950 | |
FCC | FCC PART 15 | |
75(L)*46(W)*30.5(H)mm | ||
Applications: | ||
Medical Device and ITE Equipment |
Models & Ratings | ||||||||
Model No. | Voltage (V) | Current (A) | Ripple & Noise (Max: mV) PK-PK | Voltage tolerance | Line& Load Regulation | 115 230 | Efficiency (Average) Min | Start Up delay |
UE15WCP1-XXXYYYSPA | 3.0 | 0.01-2.00 | 100 | ±10% | Y Y | 75.90% | ≤3S | |
3.3 | 0.01-2.00 | 100 | ±5% | Y Y | 75.90% | ≤3S | ||
3.6 | 0.01-2.00 | 100 | ±5% | Y Y | 75.90% | ≤3S | ||
4.0 | 0.01-2.00 | 100 | ±5% | Y Y | 75.90% | ≤3S | ||
4.2 | 0.01-2.00 | 100 | ±5% | Y Y | 75.90% | ≤3S | ||
4.5 | 0.01-2.00 | 200 | ±5% | Y Y | 75.90% | ≤3S | ||
5.0 | 0.01-2.00 | 200 | ±5% | Y Y | 75.90% | ≤3S | ||
5.2 | 0.01-2.00 | 200 | ±5% | Y Y | 75.90% | ≤3S | ||
5.5 | 0.01-2.00 | 200 | ±5% | Y Y | 75.90% | ≤3S | ||
6.0 | 0.01-2.00 | 200 | ±5% | Y Y | 75.90% | ≤3S | ||
6.0 | 0.01-2.00 | 200 | ±5% | Y Y | ≤3S | |||
6.5 | 0.01-2.00 | 200 | ±5% | Line:±2% | Y Y | ≤3S | ||
7.0 | 0.01-1.70 | 200 | ±5% | Load: ±5% | Y Y | 81.00% | ≤3S | |
7.5 | 0.01-1.50 | 200 | ±5% | Y Y | 81.00% | ≤3S | ||
8.0 | 0.01-1.50 | 200 | ±5% | Y Y | 81.00% | ≤3S | ||
8.5 | 0.01-1.40 | 200 | ±5% | Y Y | 81.00% | ≤3S | ||
9.0 | 0.01-1.40 | 200 | ±5% | Y Y | 81.00% | ≤3S | ||
10.0 | 0.01-1.40 | 200 | ±5% | Y Y | 81.00% | ≤3S | ||
10.5 | 0.01-1.30 | 200 | ±5% | Y Y | 81.00% | ≤3S | ||
12.0 | 0.01-1.25 | 200 | ±5% | Y Y | 81.00% | ≤3S | ||
13.8 | 0.01-1.05 | 200 | ±5% | Y Y | 81.00% | ≤3S | ||
15.0 | 0.01-1.00 | 200 | ±5% | Y Y | 81.00% | ≤3S | ||
16.0 | 0.01-0.85 | 200 | ±5% | Y Y | 81.00% | ≤3S | ||
17.5 | 0.01-0.80 | 200 | ±5% | Y Y | 81.00% | ≤3S | ||
18.0 | 0.01-0.80 | 200 | ±5% | Y Y | 81.30% | ≤3S | ||
19.0 | 0.01-0.75 | 200 | ±5% | Y Y | 81.30% | ≤3S | ||
20.0 | 0.01-0.75 | 200 | ±5% | Y Y | 81.30% | ≤3S | ||
22.0 | 0.01-0.65 | 200 | ±5% | Y Y | 81.30% | ≤3S | ||
23.0 | 0.01-0.65 | 200 | ±5% | Y Y | 81.30% | ≤3S | ||
24.0 | 0.01-0.63 | 200 | ±5% | Y Y | 81.30% | ≤3S |